Max. Electric Power 32 kW (100 kV, 320 mA, 20 ms / 80 kV, 400 mA, 20 ms)
  • Tube voltage Range: 40-133 kV
  • Tube current Range: 50- 400 mA
  • Maximum power: 32 kW (20 msec)
  • Nominal minimum exposure time: 1.0 msec.
  • Current-Time Range:
  • Setting Range: 0.32 – 320 mAs at 12.5% step
 
Specification of X-Ray Tube Unit (Shimadzu Tube) Model 0.7/1.3U163C-36:
  • Nominal Focal Spot Size: 0.7/1.3 mm.
  • Target Angle: 16 degrees
  • Nominal Max. Tube Voltage Radiography:133 kV
  • Max. Anode Heat Content: 300 kHU.
  • Max. Anode Heat Dissipation Rate: 1130 HU/s
  • Max. Continuous Heat Dissipation Rate: 300HU/s
  • Nominal anode input power (20 msec, 0.1msec)
  • Focus size 1.3 mm: 36.4 kW.
  • Focus size 0.7 mm: 19.9 kW.
  • Anode rotation: RPM at 70 Hz: 3800 rpm or more.
  • RPM at 60 Hz: 3200 rpm or more.